Configure FreeProbe2 Using the ProbeConfigurationTool in the FreeScan Trak Software (v3.0 or higher)

  1. Contact SHINING 3D Metrology Support and provide the serial numbers of your FreeProbe 2 and Probe Calibrator. Photos of the serial number labels are recommended.
  2. SHINING 3D Support will send you two files, as shown below.

  3. Download both files. Rename the file ending in ProbeCalibrator.asc by removing the prefix FreeScanFPB-YEABFXXXXXX_, as shown below.

  4. Copy the two files from Step 3 to the following folder:

    C:\Program Files\Shining3d\FreeTrak_TRAK\res\FreeScanTrak

  5. Locate ProbeConfigTool.exe in the folderC:\Program Files\Shining3d\FreeTrak_TRAK, then double-click it to launch the tool:

  6. Select the correct scanner model and click Connect Device. Make sure the FreeTrak system is connected via either a wired or wireless connection.

  7. Enter the FreeProbe serial number. The format must exactly match the serial number on the product label. It is recommended to copy it directly from the FreeScanFPB-YEABFXXXXXX_ProbeCalibrator.asc file.

  8. Click Config. Once the configuration is complete, a confirmation message will appear indicating that the configuration was successful.

Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.